ULTRAFLOW® 54 DN15-125
82
Kamstrup A/S · Technical description · 5512-2464_D1_GB_09.2021
8.5.5
Spare parts
Description
Type No.
Battery D-cell
1606-064
Cable retainer (secures the battery)
1650-099
2-pole plug (female)
1643-185
3-pole plug (female)
1643-187
Base PCB (66-R)
5550-517
Table 34. Spare parts for Pulse Tester.
8.5.6
Battery replacement
If Pulse Tester is used continuously, it is recommended to replace the battery once a year.
The battery connector is cut and the cable insulation is removed before the battery is connected to the terminals marked
“Batt”, the red wire to + and the black
wire to -.
Current consumption:
Current consumption with no sensors connected
400
A
Max current consumption with two ULTRAFLOW® connected
1.5 mA
<
If the base unit is fitted with battery or externally supplied, the Pulse Tester’s integral supply must be disconnected
(unplugged).
8.6
Adjustment of ULTRAFLOW® flow sensors with Kamstrup software
Adjusting of ULTRAFLOW® flow sensors is supported from Kamstrup for authorized laboratories. The adjustment is
performed through
Kamstrup’s software LabTool. Please contact the
Kamstrup Product Service in Denmark
) for additional information.
<
To prevent any attempt of fraud, LabTool is exclusively distributed to workshops/laboratories with national legal
authorization and partnership with Kamstrup.
8.7
Calibration and adjustment with NOWA
NOWA describes a standardized interface between the heat meter laboratory test equipment (e.g. flow bench) and is
mainly used in laboratories in Germany and Austria. NOWA test and adjusting of ULTRAFLOW® 54 DN15-125 (q
p
0.6…100
m³/h) is supported in combination with MULTICAL® 601/602/603. For more information about test and adjusting of
ULTRAFLOW® with NOWA, see Kamstrup documentation 5585-703 (Hardware) and 5585-706 (Software). NOWA does
not support calibration of ULTRAFLOW® 54 DN150-300 (q
p
≥
150 m
3
/h). Verification of ULTRAFLOW® 54 DN150-300 (q
p
≥
150 m
3
/h) is instead carried out via counting of pulses.