I N S TA L L AT I O N C O N T I N U E D
WARNING: Do not actuate the IAS until you have set the appropriate cracking pressure (page 11).
1.
With the probe on the ground or a table, use a 3/8" Allen wrench to loosen the Socket Cap Screws on the Locking
Collar. FIG. 3-2. This will allow the IAS™ Sampler Shaft to slide through the Seal Housing.
2.
If not already done, push the IAS™ back through the Seal Housing until the sampler probe is fully retracted inside
the Seal Housing. FIG. 3-3 and FIG. 3-4.
NOTE: On flanged models with 5 cc bite size, the sampler probe may rest flush with or slightly extend from the
end of the seal housing.
FIG. 3-4
Flanged Seal Housing
FIG. 3-3
2" MNPT Seal Housing
INSERTABLE/RETRACTABLE MODELS INSTALLATION
3.
Re-tighten the Socket Cap Screws on the Locking Collar. This will prevent the IAS™ shaft from sliding and the
probe from getting damaged during mounting.
Seal Housing
Seal Housing