Full Digital Ultrasonic Diagnostic Instruments
User’s Manual V1.00
- 16 -
Menu
】
;
2. Press
『
Set
』
or
『
Back
』
key to realize compression curve processing. The compression curve falls
into eight levels, which are 0-7 and displayed in the
〖
Comp Curve
〗
of the left menu.
Note: If use 3.5 MHz probe, it is recommended to choose 5 or 6; if use 5.0 MHz probe,
recommend choose 3 or 4; if use 6.5 MHz probe, recommend choose 1 or 2; if use 7.5
MHz probe, recommend choose 0 or 1.
7.6.9 Image post-process adjustment
1. In real-time or frozen mode, move the cursor to the submenu
〖
IP
〗
of
【
B Mode Menu
】
;
2. Press
『
Set
』
or
『
Back
』
key to realize image post-process. Image post-process falls into four
levels, which are 0-3 and displayed in the
〖
IP
〗
of the left menu. NO.2 is recommended when
normal use.
7.6.10 Image optimization
1. In real-time mode, move the cursor to the submenu
〖
iImage
〗
of
【
B Mode Menu
】
;
2. Press
『
Set
』
or
『
Back
』
key to realize 10 kinds of image optimization processing. Image
optimization falls into 0-9 and displayed in the
〖
iImage
〗
of the left menu. The default value is 4.
7.6.11 Depth range selection
In real-time mode, rotate
『
Function
』
knob on the panel to get images in different depth,
B, B/B,
4B mode has 10 kinds of depth range; B/M, M mode has 6 kinds of depth range. Users also can fast
select depth range with 0-9 number keys.
7.6.12 Image up/down reverse
1. In real-time or frozen mode, move the cursor to the submenu
〖
V Rev
〗
of
【
B Mode Menu
】
;
2. Press
『
Set
』
or
『
Back
』
key to realize image vertical reversal.
7.6.13 Image left/right reverse
1. In real-time mode, move the cursor to the submenu
〖
H Rev
〗
of
【
B Mode Menu
】
;
2. Press
『
Set
』
or
『
Back
』
key to realize image horizontal reversal that change the probe scanning
direction.
7.6.14 Angle/Width change
1. In real-time mode, move the cursor to the submenu
〖
Angle Change
〗
of
【
B Mode Menu
】
in B
mode;
2. Press
『
Set
』
or
『
Back
』
key to convert the scanning angle/width. Angle/width change falls into
0, 1, 2, 3 and displayed in the
〖
Angle Change
〗
of the left menu. (Convex array for scanning
angle and linear array for scanning width)
7.6.15 Color selection
1. In real-time or frozen mode, move the cursor to the submenu
〖
B Color
〗
of
【
B Mode Menu
】
.
2. Press
『
Set
』
or
『
Back
』
key to achieve the conversion of eight kinds of colors (including one kind
of black and white).
7.6.16 Image freeze/unfreeze
In real-time mode, press
key to freeze the image; in frozen status, press
key to unfreeze
the image.
7.7 Puncture guide
In real-time B mode and in the status of
【
B Mode Menu
】
, press
key, two puncture guide lines
will be showed on the screen, press direction keys
to change the angle of the first puncture guide
line, press direction keys
to change the start position of the first puncture guide line.