Full Digital Ultrasonic Diagnostic Instruments
User’s Manual V1.00
- 9 -
5.3.2 Ultrasonic probe disconnection
Shutdown the system; rotate the locking knob 90 degree counterclockwise, pull out the probe
connector vertically.
5.4 Install/Remove the battery
1. Install the battery
Shutdown the system; remove the battery cover on the right side of main unit. According to the
direction of battery indication to install the battery and then install the battery cover.
2. Remove the battery
Shutdown the system; remove the battery cover on the right side of main unit, push forward
the battery putter at the bottom of main unit and the battery is launched, remove the battery.
Fig. Install, Remove the battery
5.5 Install foot switch
Shutdown the system; insert the plug of foot-switch into the “
” socket at the back of the main
unit.
5.6 Connection to video recorder
1. Shutdown the system; connect the equipotential terminal ( ) of the video recorder to the
earthing;
2. Connect one end of the video cable to the video recorder and the other end to the video output
interface on the rear panel of the main unit;
3. Insert one end of power plug (jack) of the video recorder to its power input socket, the other
end to the power supply socket.
5.7 Connection to power
1. Connect to the power adapter
Insert the output plug of adapter into DC power input port, which is on the right of main unit.
2. Connect to the main power supply
Insert the power plug (jack) furnished with the machine into power input socket of the power
adapter, the other end to the mains socket-outlet. The instrument uses three-core power supply.
It connects with the protective earth line when power plug inserts into its socket.
Warning:
1. Adapter has no switch. The isolation of the system with the MAINS used to unplug the
adapter as the intended isolation means.
2. The equipment should be used only with adapter provided by Kaixin Company.
3. To avoid damaging power adapter or harming people by unexpected fallen, make sure
the power adapter is placed on the leveled desk.
4. The operator must not touch signal input/ signal output and patient simultaneously.
Battery cover
Battery putter
Bottom of the
main unit
Battery cover
Attention:
The waterproof grade of foot switch is IPX1.