31
considered. If the measured field strength in the location in which the
JPD-FR410 is used exceeds the applicable RF compliance level above,
the JPD-FR410 should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the JPD-FR410.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should
be less than 3 V/m.
Recommended separation distances between portable and mobile RF
communications equipment and the EQUIPMENT or SYSTEM -for
EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
The Infrared Thermometer JPD-FR410 is intended for use in an
electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Infrared Thermometer
JPD-FR410 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communications
equipment
(transmitters)
and
the
Infrared
Thermometer JPD-FR410 as recommended below, according to the
maximum output power of the communications equipment.
Rated
maximum
output
power
of
transmitter
W
Separation distance according to frequency of
transmitter
m
80 MHz to 800 MHz
800 MHz to 2,5 GHz