28
RF
emissions
CISPR 11
Class B
The Infrared Thermometer JPD-FR410
is suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Guidance and manufacturer's declaration – Electromagnetic
immunity –for all equipment and systems
Guidance and manufacturer´s declaration – Electromagnetic immunity
The Infrared Thermometer JPD-FR410 is intended for use in the
electromagnetic environment specified below. The customer or the user
of the Infrared Thermometer JPD-FR410 should assure that it is used
in such an environment.
Immunity
test
IEC
60601
test
level
Compliance
level
Electromagnetic
environment- guidance
Electrostatic
discharge
(ESD)
IEC
61000-4-2
±6kV
contact
±8 kV
air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Power
frequency
(50/60 Hz)
magnetic
field
IEC
61000-4-8
3 A/m
3 A/m
Power frequency
magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.