9 Startup
32
9.4
Window opening function
The window opening function can be set using three parameters. The principle switching points are de-
fined using the switching point and release point. In addition, hysteresis can also be set for the switching
point and release point. The hysteresis function is deactivated as soon as the hysteresis
parameter >= (Sp - Rsp)/2 is configured.
9.5
Window closing function
The window closing function constitutes the inverted function of the window opening function.
(1)
Release point
(2)
Switching point
(3)
Flow velocity
(4)
Hysteresis
(1)
(2)
(3)
[
cm/s
]
(4)
(4)
0
1
(1)
Release point
(2)
Switching point
(3)
Flow velocity
(4)
Hysteresis
(1)
(2)
(3)
[
cm/s
]
(4)
(4)
0
1
Summary of Contents for PINOS L02
Page 1: ...40604100T90Z000K000 V2 00 DE EN 00691112 JUMO PINOS L02 Betriebsanleitung Operating Manual...
Page 2: ......
Page 4: ......
Page 18: ...3 Ger teausf hrung identifizieren 16...
Page 28: ...7 Anzeige und Bedienelemente 26...
Page 42: ...10 Wartung 40...
Page 44: ...11 China RoHS 42...
Page 45: ......
Page 47: ...40604100T90Z000K000 EN 00691112 JUMO PINOS L02 Calorimetric flow sensor Operating Manual...
Page 48: ......
Page 52: ...1 Safety information 6...
Page 62: ...3 Identifying the device version 16...
Page 86: ...10 Maintenance 40...
Page 88: ...11 China RoHS 42...
Page 89: ......
Page 91: ......
Page 92: ......