9 Adjustment options
40
Characteristics
Instrument
version
Display format: xx.yy
xx = hardware version
yy = software version
Densitiy
correction
Density correction is important for the hydrostatic level measurement for
media with densities other than 1 (e.g. pure water).
Possible range setting: 0.100—5.000 (kg/dm
3
).
Example
Range: 0—1 bar
Scaling:
0—100% => 0—5000 l
Unit of
temperature
display
The display unit of the measured sensor temperature can be selected as:
❏
°C
❏
°F
100
80
60
40
20
0
20
10
0
0
20
40
60
80
0
1
2
3
4
100
Output current
[%]
Pressure [%]
(b)
(a)
20
10
15
5
0
0
1
2
3
4
(a)
(b)
(c)
Starting point
Starting point
linear
15
5
(a) = Lin
= linear
(b) = SLin = square root, linear up to
starting point
(c) = SoFF = square root, switched off
up to starting point
Medium Density
Measured
pressure
Output current
Display LCD
Water
1.0
0.5 bar
12 mA
2500 liters
Gasoline
0.8
0.5 bar
14 mA
3125 liters
All manuals and user guides at all-guides.com
Summary of Contents for dTRANS p02
Page 2: ...All manuals and user guides at all guides com...
Page 5: ...All manuals and user guides at all guides com...
Page 47: ...45 12 Maintenance All manuals and user guides at all guides com...
Page 50: ...13 Appendix 48 All manuals and user guides at all guides com...
Page 52: ...13 Appendix 50 All manuals and user guides at all guides com...
Page 56: ...13 Appendix 54 All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 61: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Page 62: ...All manuals and user guides at all guides com...
Page 63: ...All manuals and user guides at all guides com...