
8 Calibration
78
The zero point that was determined by the instrument is shown.
✱
Accept the value with the
key or
use the
key to reject the value.
The instrument returns to the measurement mode.
8.3.2
2-point calibration
Requirements
- The supply voltage for the instrument must be present.
see Chapter 4 “Electrical connection”, page 12ff.
- A combination electrode must be connected to the transmitter.
- “REDOX” (ORP) must be configured for the sensor and “CUSTOMIZED” or
“%” for the UNIT in the basic settings.
- Calibration must be enabled,
see Chapter 6.9.1 “Administrator levels”, page 39.
- The transmitter is in the measurement mode.
✱
Immerse the combination electrode in a buffer solution with a known ORP.
✱
Start the calibration (by pressing the
key, or via the Administrator level).
PGM
EXIT
The ORP of the sample solution is not dependent on temperature!
CAL
All manuals and user guides at all-guides.com