4 Electrical connection
18
Symmetrical connection of a combination electrode with separate
temperature sensor
✱
Lead the connecting cables in through the cable glands.
✱
Lay the signal cables as shown in the diagram. Use the cable clips (3) to
clamp each signal cable to the shielding.
✱
Connect the cores as assigned below, and see Chapter 4.5 “Terminal
assignments”, page 19.
✱
Push the plug-in terminals for row 1 (1) and row 2 (2) into the sockets in the
instrument.
Sensor
connection
(1)
(2)
(3)
pH combination electrode
SC = shielding (screen)
Row 1
Row 2
Separate temperature sensor
Ground pin or conductive pipe/
(1)
(2)
(3)
(4)
?
reservoir wall at measurement point
Triaxial cable
! remove plug-in link 3-5 !
In environments with difficult EMC conditions, a coaxial cable with a double
shielding must be used. A shielded 2-core cable is required for connecting the
temperature probe.