83
27 Controller
27.1 Configuration
Possible
combinations
The control functions of outputs 1 and 2 can be freely combined
1
:
- Controller off
- Limit controller
- Pulse width controller
- Pulse frequency controller
1
Exception: When using a modulating controller, outputs 1 and 2 must
have the same configuration.
The controller functions are determined by the following parameters:
1
Chapter 26.4 “Controller options - C211”, page 76 or
Chapter 26.5 “Controller outputs - C212”, page 77 or
Chapter 26.6 “Other outputs I - C213”, page 78 or
Chapter 26.7 “Other outputs II - C214”, page 79.
2
Chapter 20 “Parameter level of the conductivity indicator”, page 59ff.
3
Chapter 19 “Operator level of the conductivity indicator”, page 58ff.
For an explanation of the terminology used,
Chapter 33 “Glossary”, page 93ff.
Configuration level
1
Parameter level
2
Operating level
3
C211
C212
C212
C213
C214
Controller off
– –
– –
– –
– –
– –
– –
Limit
controller
MIN / MAX
contact
make / break
contact
– –
– –
Switching differential
HYS
Pull-in delay
Ond
Drop-out delay
Ofd
Setpoint
SP(r)
Pulse width
controller
MIN / MAX
contact
make / break
contact
– –
– –
Proportional band
Pb
Derivative time
dt
Reset time
rt
Minimum ON time
tr
Pulse period
CY
Output level limit
Y1
or
Y2
Setpoint
SP(r)
Pulse frequency
controller
MIN / MAX
contact
make / break
contact
– –
– –
Proportional band
Pb
Derivative time
dt
Reset time
rt
Minimum pulse width
tr
Maximum pulse frequency
Fr
Output level limit
Y1
or
Y2
Setpoint
SP(r)
Modulating
controller
MIN / MAX
contact
make / break
contact
– –
– –
Proportional band
Pb
Derivative time
dt
Reset time
rt
Minimum ON time
tr
Pulse period
CY
Output level limit
Y1
or
Y2
Actuator time
tt
Setpoint
SP(r)
Proportional
controller
MIN / MAX
contact
make / break
contact
Proportional
controller 1
Proportional
controller 2
Proportional band
Pb
Derivative time
dt
Reset time
rt
Output level limit
Y1
or
Y2
Setpoint
SP(r)