58
19 Operator level of the conductivity indicator
19.1 Settings
Pre-
conditions
How to access the operating level, or leave this level,
Chapter 9.2 “Principle of operation”, page 24ff.
The operating level must be unlocked,
“Unlocking the levels”, page 27, (code word 0110)
.
1
depending on the configured measurement range,
“rAnG - CELL - ALPH”, page 62.
It is possible that not all of the following parameters are needed or
displayed, depending on the configuration of the controller
functions.
For an explanation of the terminology used,
Chapter 33 “Glossary”, page 93ff.
How to configure controllers,
Chapter 27.1 “Configuration”, page 83ff.
Designation
Parameter
(display)
Value range
Factory
setting
displayed if ... is
configured
see
Configura-
tion
parameter
Setpoint 1
SP(r)1
0 — 0.5
µ
S
to
0 — 200mS
1
0.00
K1
C211
Setpoint 2
SP(r)2
1.00
K2
Setpoint 3
SP(r)3
-0.00
Setpoint changeover
C112
Setpoint 4
SP(r)4
1.00
Code word
CodE
4-digit
0000
Limit LK A (K1)
SP A
corresponds to
“rAnG”
or
-50.0 to 250.0°C
-1.00
K1
C214
Limit LK b (K2)
SP b
K2
Limit LK C (K3)
SP C
K3
C113
Limit LK d (K4)
SP d
K4
Limit LK E (K5)
SP E
K5
C114
Process value
input 2
(temperature)
InP2
(°C)
25
C111
Alarm tolerance
AL1
corresponds to
“rAnG”
0
Controller alarm
messages
C211 or
C213
Alarm delay
AL2
0 to 9999 sec
300