
20 Parameter level of the conductivity indicator
60
Switching differential,
controller 1
HYS1
0001 to
9999
2% of full
scale
Relay 1, limit value
in C211
Switching differential,
controller 2
HYS2
Relay 2, limit value
in C211
Switching differential,
controller 3
HYS3
Relay 3, limit value
in C213
Switching differential,
controller 4
HYS4
Relay 4, limit value
in C213
Switching differential,
controller 5
HYS5
Relay 5, limit value
in C214
Pull-in delay 1
Ond1
0.00 to
999.9 sec
1.0
Relay 1, limit value
in C211
Pull-in delay 2
Ond2
Relay 2, limit value
in C211
Pull-in delay 3
Ond3
Relay 3, limit value
in C213
Pull-in delay 4
Ond4
Relay 4, limit value
in C213
Pull-in delay 5
Ond5
Relay 5, limit value
in C214
Drop-out delay 1
Ofd1
0.2 sec
Relay 1, limit value
in C211
Drop-out delay 2
Ofd2
Relay 2, limit value
in C211
Drop-out delay 3
Ofd3
Relay 3, limit value
in C213
Drop-out delay 4
Ofd4
Relay 4, limit value
in C213
Drop-out delay 5
Ofd5
Relay 5, limit value
in C214
Maximum pulse
frequency 1
Fr1
0 to 150
pulse/min
100
Relay 1, pulse frequency
in C211
Maximum pulse
frequency 2
Fr2
Relay 2, pulse frequency
in C211
Pulse period 1
CY1
1.0 to
999.9 sec
20.0
Relay 1, pulse width
in C211
Pulse period 2
CY2
Relay 2, pulse width
in C211
Output level limit
Relay 1
Y1
0 to 100% 100
Relay 1, pulse frequency or pulse
width
in C211
Output level limit
Relay 2
Y2
Relay 2, pulse frequency or pulse
width
in C211
Filter constant
dF
0 to 100
sec
0.6
Actuator time
tt
15 to 3000
sec
60
Modulating controller
in C211
Parameter
Display
Value
range
Factory
setting
displayed if ... is configured