18 Conductivity indicator
50
Chapter 21.2 “Analog inputs - C111”, page 61.
The instrument is in the measurement mode,
“Operating modes and states”, page 24.
Procedure
The upper display shows the compensated conductivity value of the solution
being measured.
The indicated conductivity depends on the manually set temperature, see
Temperature setting
,
below
and the set (or automatically acquired) temperature coefficient (TC),
Chapter 18.8.1 “Automatic determination of the temperature coefficient,
using manual temperature entry”, page 55.
The lower display shows the manually entered temperature setting.
18.4 Manual temperature entry
Initial condition
The temperature acquisition is configured as
“Manual temperature compensation”,
Chapter 21.2 “Analog inputs - C111”, page 61.
The operating level is unlocked,
“Unlocking the levels”, page 27, (code word 0110)
The instrument is in the measurement mode,
“Operating modes and states”, page 24.
Procedure
✱
Press
briefly and repeatedly, until “InP2” is displayed.
Use the
and
keys to set the temperature that is shown
✱
Press the
key (confirmation)
✱
Press the
key (to return to measurement mode) or cancel the entry
18.5 Measurement with automatic temperature compensation
Initial condition
The temperature acquisition has been configured as
“Automatic temperature compensation with Pt100 or Pt1000”,
Chapter 21.2 “Analog inputs - C111”, page 61.
The instrument is in the measurement mode,
“Operating modes and states”, page 24.
Procedure
The temperature measurement for the medium cannot be altered manually.
PGM
PGM
EXIT