Maintenance
25
1. Clean the surface of the device with a damp cloth.
Use a little dish soap for cleaning. If in doubt, ask Technical Service about
alternative cleaning agents.
2.
Remove residue in the bath using a diluted soap-based
cleaning
solution. To do so, fill the device with your diluted cleaning
solution and
temper to 60°C for several hours. Then empty the bath and
wipe it with
a cloth.
3.
Regularly remove limescale deposits on the bathtub bottom using a
limescale cleaner.
The device has now been cleaned.
9.4 Device storage
Take your device out of operation if you have not used it for a long time or, for
example, it is to be sent to Technical Service for repair. Follow the procedure
described to ensure that your device continues to function reliably even after
being stored for a long period.
The device is switched off.
1. Empty the water bath completely.
2. Disconnect the device from the power supply.
3. Clean the inside and outside of the water bath.
4. Close the drain valve (not for PURA 4).
5. Store the device in a dust-free, dry and frost-free location.
The water bath is taken out of operation and stored away. It can be put into
operation again as needed.
Summary of Contents for Pura 10