7 Initial operation
7.1 Connect the device to the power supply
This section describes how to connect the device electrically using the mains
plug.
The device is at its installation location.
1. Insert the device power cable plug into the mains connection socket on the
rear of the device.
2. Insert the mains cable plug into the power supply socket.
The device is connected to the electricity.
7.2 Required water quality
Only water is permitted as a bath fluid. Follow the specifications on water
quality to ensure long-term operation without issue.
NOTE
Risk of corrosion due to unauthorized bath fluid!
Unsuitable bath fluids not approved by JULABO can damage the device.
• Only water is permitted as a bath fluid
• JULABO recommends a water mixture of 70% softened/descaled water
and 30% tap water
• Consult JULABO before using a bath fluid other than what is
recommended
• Do not use water containing iron Corrosion also possible with stainless
steel.
• Do not use water containing chlorine Pitting corrosion possible.
• Do not use distilled or deionised water Corrosion also possible with
stainless steel.
• Do not use water containing calcium carbonate A high lime content leads to
limescale formation in the bath.