14
5.3 Technical data
Temperature stability measured in accordance with DIN 12876 with water at
70°C. Performance specifications apply at an ambient temperature of 20°C.
Grouping of the device acc. to CISPR 11:
• The device is an ISM device of group 1, which uses high frequency for
internal purposes
• Class A: Use in an industrial electromagnetic environment
In accordance with IEC 61010-1, the device is designed for safe operation under
the following ambient conditions:
• Indoor use
• Altitude up to 2000 m above sea level
• Ambient tempe5 … +40°C
• Maximum relative humidity 80% for temperatures up to 31°C, decreasing
linearly down to 50% relative humidity at 40°C
• Mains voltage fluctuations up to ±10% of the nominal voltage permissible
if not otherwise specified
• Contamination level 2
Classified in accordance with DIN 12876-1:
• Class I
Protection class according to EN 60 529:
• Protection class IP21
Technical data
PURA
4/10/14/22/30
Working temperature range*
°C
+18 … +99.9
Overheating protection, fixed
setting
°C
130
Temperature stability**
°C
±0.15
Temperature resolution
°C
0.1
Temperature setting
Digital
Temperature control
PID1
Display
LED
Bath tank, basic material
Stainless steel
*with counter-cooling and bath cover (accessories)
* with bath cover (accessory)