![Julabo Pura 10 Original Operating Manual Download Page 20](http://html1.mh-extra.com/html/julabo/pura-10/pura-10_original-operating-manual_2030684020.webp)
Operation
20
NOTE
Condensate forms from water vapor!
The formation of water vapor during operation can cause condensation to
build up in the immediate vicinity.
• Ensure sufficient distance between the water bath and adjacent
equipment
• When using the bath cover (accessory): Do not close vent holes
NOTE
Oxidation from insert racks or non-ferrous metal samples!
Insert racks or non-ferrous metal samples can cause electrochemical
oxidation and damage the bath tank.
• Do not use insert racks and samples made of non-ferrous metal
• Only use original JULABO insert racks
NOTE
Falling bath fluid level!
In the case of a prolonged temperature control application, the level of bath
fluid in the bath tank can fall below the alarm limit due to evaporation. A low
liquid level alarm is triggered and temperature control application stops.
• In the event of a prolonged temperature control application, regularly
check the fill level of bath fluid in the bath tank
• If the level is low, refill the bath fluid
The device is ready for use.
1. Switch the device on at the mains switch.
2. Press
[OK]
.
The actual temperature is displayed.
3. Set the desired setpoint temperature using the arrow keys.
4. Press
[MENU]
.
The setpoint temperature is saved. The display flashes briefly. The device
starts the temperature control application at once. An acoustic signal
sounds at 0.5°C before the setpoint temperature. The temperature control
application can be stopped with the
[OK]
key.
For temperature control applications near or below the ambient
temperature: Use a cooling coil or JULABO immersion cooler.
Summary of Contents for Pura 10
Page 1: ...Water baths Original operating manual 1 950 6080 us V03 02 2022 ...
Page 31: ......
Page 32: ......