LC4 F
19
The unit also enters the safe operating state "OFF" or "r
OFF" after a mains power interruptance. The temperature
values entered via the keypad remain in memory. With the
controller in keypad control mode, press the start/stop key to
restart operation.
With the controller in remote control mode, the personal
computer must first resend the parameters set via the
interface before operation of the controller may be restarted.
5.7. Setting the temperature
A
Setting the working temperature:
1. Press the setpoint key
.
The indicator light blinks and the effective value appears
on the MULTI-DISPLAY (LED).
2. Use the cursor keys
to move left or right on
the display until the numeral you wish to change is
blinking.
3. Use the increase/decrease arrows
to change
the selected numeral (-, 0, 1, 2, 3, ... 9).
4. Press enter
to store the selected value (example:
37.0 °C).
The working temperature is maintained constant after a short
heat-up time.