Menu functions
26
6. Menu
functions
A
S
The setting of the parameters and the setpoints of the
controller can be dialed up at any time through the menu in
the working area and safety area of the keypad.
The ATC menu option, which is in the safety area of the
keypad, can only be used for calibrating the safety
sensor.
•
Enter or exit the menu level by pressing the left arrow
and enter
at the same time.
•
Using the cursor keys
select the menu
functions one by one.
Example:
ATC
AO
EP
EPH ......
6.1. Sensor calibration - ATC / SF or AF-
Laborregler (T
T
)
Eichbad (T
M
)
The ATC (Absolute Temperature Calibration) function serves to
calibrate the working sensor „AF“ and the safety sensor „SF“.
Example: calibrate the working sensor „AF“
•
Place the working sensor "AF" and the temperature sensor of
a thermometer in the calibration bath..
•
Determine the difference temperature (
Δ
T=TM - TT)
and store as correcting factor (example
Δ
T = -0.26 °C) as
follows:
•
With the edit keys
change the menu function from
"Atc 0" to "Atc 1" and then press enter
.
•
Using the cursor keys
and the edit keys
set the correcting factor (example -0.26 °C) and then
press enter
.