deflate when pressures reaches 40 kPa (300 mmHg), detach the cuff from the arm and press the START/
STOP button to stop inflation.
Before use, make sure the device functions safely and is in proper working condition. Check the device,
do not use the device if it is damaged in any way. The continuous use of a damaged unit may cause injury,
improper results, or serious danger.
Do not wash the cuff in a washing machine or dishwasher!
The service life of the cuff may vary by the frequency of washing, skin condition, and storage state.
The typical service life is 10000 times.
It is recommended that the performance should be checked every 2 years and after maintenance and
repair, by retesting at least the requirements in limits of the error of the cuff pressure indication and air
leakage (testing at least at 50mmHg and 200mmHg).
Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the local
guidelines.
Manufacturer will make available on request circuit diagrams, component part lists, descriptions,
calibration instructions, etc., to assist to service personnel in parts repair.
The plug/adapter plug pins insulates the device from the main supply. Do not position the device in a
position where it is difficult to disconnect from the supply mains to safely terminate operation of ME
equipment.
The operator shall not touch output of batteries /adapter and the patient simultaneously.
Cleaning :
Dust environment may affect the performance of the unit. Please use the soft cloth to clean
the whole unit before and after use. Don’t use any abrasive or volatile cleaners.
The device doesn’t need to be calibrated within two years of reliable service.
If you have any problems with this device, such as setting up, maintaining or using, please contact the
SERVICE PERSONNEL of Joycare. Don’t open or repair the device by yourself in the event of malfunctions.
The device must only be serviced, repaired and opened by individuals at authorized sales/service centres.
Please report to Joycare if any unexpected operation or events occur.
Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing of small
parts. It is dangerous or even fatal.
Be careful to strangulation due to cables and hoses, particularly due to excessive length.
At least 30 min required for ME equipment to warm from the minimum storage temperature between
uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the
maximum storage temperature between uses until it is ready for intended use.
This equipment needs to be installed and put into service in accordance with the information provided
in the ACCOMPANYING DOCUMENTS;
Wireless communications equipment such as wireless home network devices, mobile phones, cordless
telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least
a distance d away from the equipment. The distance d is calculated by the MANUFACTURER from the 80
MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate.
Please use ACCESSORIES and detachable parts specified/ authorised by MANUFACTURE.
Otherwise, it may cause damage to the unit or danger to the user/patients.
There is no luer lock connectors are used in the construction of tubing, there is a possibility that they
might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood
vessel.