47
Recommended separation distances between portable and mobile
RF communications equipment and the Model MH-DR103
The model JC-114N (MH-DR103) is intended for use in the electromagnetic environment in which radiated Rf disturbances are controlled. The
customer or the user of the model JC-114N (MH-DR103) can help prevent electromagnetic inference by maintaining a minimum distance between
portable and mobile Rf communications equipment (transmitters) and the model JC-114N (MH-DR103) as recommended below , according to the
maximum output power of the communications equipment.
RaTeD MaxIMuM
ouTPuT PoWeR of
TRaNsMITTeR W
sePaRaTIoN DIsTaNCe aCCoRDING To fReQueNCy of TRaNsMITTeR M
150kHz To80MHz
D=1.2√P
80MHz To 800MHz
D=1.2√P
800MHz To 2.5GHz D=2.3√P
0.01
0.12
0.23
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
2.3
for transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters(m) can be estimated using the
equation applicable to the frequency of transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1 at 80MHz and 800MHz, the separation distance for higher frequency range applies.
Note 2 These guidelines may not apply in all situations. electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.