31 DePuy Synthes Instructions for Use Air Pen Drive
4.
Rinse with tap water.
Rinse device with cold tap
water for a minimum of 2 minutes. Use a syringe or
pipette to flush lumens and channels.
5.
Clean with detergent.
Clean the device manually
under running warm water using an enzymatic
cleaner or detergent for a minimum of 5 minutes.
Manipulate all moving parts under running water.
Use a soft-bristled brush and/or soft lint-free cloth to
remove all visible soil and debris.
Follow the enzymatic cleaner or detergent manufac-
turer’s instructions for use for the correct tempera-
ture, water quality and concentration/dilution.
6.
Rinse with tap water.
Rinse the device thoroughly
using cool to lukewarm running water for a mini-
mum of 2 minutes. Use a syringe, pipette or water
jet to flush lumens and channels. Actuate joints,
handles and other movable device features in order
to rinse thoroughly under running water.
7.
Visually inspect device.
Inspect the cannulations,
coupling sleeves, etc. for visible soil. Repeat steps
1–6 until no visible soil remains.
8.
Load washing basket.
Place devices in the specially
designed tray for machine washing supplied by
Synthes (68.001.800) as shown on the next page or
refer to the loading plan (DSEM/PWT/1116/0126).
68.001.800
Care and Maintenance
Cleaning and Disinfection
Automated Cleaning Instructions with Manual Pre-cleaning