28 DePuy Synthes Instructions for Use Air Pen Drive
Follow the enzymatic cleaner or detergent manufac-
turer’s instructions for use for correct temperature,
water quality and concentration/dilution.
6.
Rinse with tap water.
Rinse the device thoroughly
using cool to lukewarm running water for a mini-
mum of 2 minutes. Use a syringe, pipette or water
jet to flush lumens and channels. Actuate joints,
handles and other movable device features in order
to rinse thoroughly under running water.
7.
Wipe/Spray disinfection.
Wipe off or spray the
surfaces of the devices with a minimum of 70%
alcohol-based disinfectant.
8.
Visually inspect device.
Inspect the cannulations,
coupling sleeves, etc. for visible soil. Repeat steps
1–8 until no visible soil remains.
9.
Final rinse with de-ionized/purified water.
Final rinse with de-ionized or purified water for a
minimum of 2 minutes.
10.
Dry.
Dry device using a soft lint-free cloth or medical
grade compressed air. If smaller devices or cannula-
tions contain residual water, blow dry with medical
grade compressed air.
Care and Maintenance
Cleaning and Disinfection
Manual Cleaning Instructions