Chapter 6
Scope
Configuring the P5000 Scope
User Manual
790000002/00
37
Configuring the P5000 Scope
Scope
connection
1
Plug in your JDSU scope into a USB port from the 2000 Platform.
2
Push the button
H
OME
3
Validate the
Scope
function
.
4
Connect probe with the fiber being inspected.
You may select this option while other options are already selected (e.g.
OTDR).
Configuring the
Scope
1
Press
Setup
key to configure the test.
The following screen displays:
Test
On the line
Profile
, select the Profile which will be used for the test of
fiber connector:
–
IEC SM-UPC: Pass/Fail criteria for single-mode UPC connectors
from IEC 61300-3-35 standard.
–
IEC SM-APC: Pass/Fail criteria for single-mode APC connectors
from IEC 61300-3-35 standard.
Profiles contain the analysis parameters by which PASS/FAIL criteria are
determined.
Once the line is selected, you can also add a new profile, clicking on the
Add
"Adding a new profile" page 38
).
Fig. 18
P5000 Scope Setup
2000 Platform English rev00.book Page 37 Lundi, 20. juin 2011 1:47 13