98
UMG 96RM-M
Adress
Format
RD/WR
Unit
Note
Adjustment Range Default
81
SHORT
RD/WR
-
Secondary address,
0..99
extended section 1
82
SHORT
RD/WR
-
Secondary address,
0..99
extended section 2
83
SHORT
RD/WR
-
Secondary address,
0..99
extended section 3
84
SHORT
RD/WR
-
Secondary address,
0..99
extended section 4
500
SHORT
RD/WR
-
Terminal assignment, I L1
-3..0..+3
1)
+1
501
SHORT
RD/WR
-
Terminal assignment, I L2
-3..0..+3
1)
+2
502
SHORT
RD/WR
-
Terminal assignment, I L3
-3..0..+3
1)
+3
503
SHORT
RD/WR
-
Terminal assignment, U L1
0..3
1)
1
504
SHORT
RD/WR
-
Terminal assignment, U L2
0..3
1)
2
505
SHORT
RD/WR
-
Terminal assignment, U L3
0..3
1)
3
506
SHORT
RD/WR
-
Clear min. and max. values
0..1
0
507
SHORT
RD/WR
-
Clear energy meter
0..1
0
508
SHORT
RD/WR
-
Force write EEPROM
0..1
0
Note: Energy values and minimum and maximum values are written to the EEPROM every 5 minutes.
509
SHORT
RD/WR
-
Voltage connection diagram
0..8
2)
0
510
SHORT
RD/WR
-
Current connection diagram
0..8
0
511
SHORT
RD/WR
-
Relative voltage for
THD and FFT
0, 1
0
The voltages for THD and FFT can be shown on the display as L-N or L-L values. 0=LN, 1=LL
600
UINT
RD/WR
-
Metering range exceedance
0..0xFFFFFFFF
746
SHORT
RD/WR
s
Period of time after which the
backlight will switch to standby
60 .. 9999
900
747
SHORT
RD/WR
s
Brightness of the standby backlight
0 .. 9
0
1)
0 = No measurement of the current or voltage path.
2)
The setting 8 is equal setting 0.