8
Press the key
to
finish setting and return to
normal weighing mode.
Part VII Error Message & Simple Troubleshooting
Error
message
Cause
Solutions
Err
Beyond the display range
Reduce the
weight
Err2
Initial zero point exceeds +30% (take 10% as
reference basis), 1. There is something on the pan;
2. There is interference from other articles; 3.
LOAD CELL failed.
1. Take away the
articles that result
in interference. 2.
Replace LOAD
CELL or contact
service
department.
-Err2
Initial zero point exceeds –30% (take 10% as
reference basis), 1. There is something on the pan;
2. There is interference from other articles; 3.
LOAD CELL failed.
1. Take away the
articles that result
in interference. 2.
Replace LOAD
CELL or contact
service
department.
Err3
Higher than A/D resolution range – FFFFF, 1. A/D
failed; 2. LOAD CELL failed
Replace LOAD
CELL or contact
service
department
-Err3
Lower than A/D resolution range – XXXXX, 1. A/D
failed, 2. LOAD CELL failed
Replace LOAD
CELL or contact
service
department
Err4
EEPROM Checksum error
If the error
message appears
once again after
the scale is
powered on
again, please
re-weld EEPROM
or contact service
department.
Err5
The weighed articles are overload (max. weighing
capacity: +9e)
Immediately take
away the
overload articles
from the pan and
PAGE
JADEVER
http:www.jadever.com.tw
24