Part I Preparations before Usage
1-1. Precautions before Usage
◎
Please charge the battery before using for the first time.
Charging instruction: when the battery symbol appears on the upper left of the screen, it
indicates that the power nearly runs out. Please recharge as soon as possible. Once
the indicator turns green from red, it indicates that the recharging is completed, which
takes about 8 hours. Avoid using before recharging is completed.
◎
Be sure to select proper power supply before using, where a separate power outlet will
help avoid being interfered.
Select a flat and
◎
stable
installation site.
◎
Avoid an operation environment where the scale will be exposed to strong wind,
vibration and electromagnetic wave.
(e.g. air inlet of air conditioner or fan, or near heavy-duty
machines).
Please select an environment where the temperature is maintained within 0
◎
℃
~ 40
,
℃
avoid drastic temperature change.
15
◎
-minute preheating is required before usage.
Please select 110~240 VAC±10% 50/60HZ,6V4Ah BATTERY. Improper battery or
※
connection may cause system damage or other adverse effects.
1-2. Specifications and Characteristics
Model
JWI600
Input Sensitivity
0.13uV/DIV
Input Voltage Range
-0.5mV to 16.5mV
Supply Voltage of Load
Cell
5V DC ±5%
Driving Capacity of Load
Cell
Up to 8 untis of load cell at 350 ohm (120mA)
Non-linearity
0.006% of full capacity
PAGE
JADEVER
http:www.jadever.com.tw
2