V03
JCAII
7
6) Press key to save the settings and return to weighing mode.
7
)
Press key to change its unit: kg, lb, g.
7-5 Calibration
= remote platform
= main scale
Note:
z
Please calibrate the main scale first.
z
The calibration steps of main scale and remote platform are the same. Press key to
switch the calibration for the main scale or remote platform before the zero point calibration.
z
You can only do the calibration for the remote platform after you have set the max. capacity
and division of the remote platform.
Here take JCAII-6K as an example.
1. Press key
and hold
while turning on the scale. Do not release
till
is displayed in the Unit Weight window.
2. With no load on the weighing pan, press key to start the zero point calibration and wait
till
is displayed in the Unit Weight window. Input the first calibration weight value by the
numeric key (6K model, 1/3 of full load is 2kg). And then put the corresponding weights on.
3... Press key and wait till
is displayed in the Unit Weight window. And then
put the corresponding weights on.
4. Press key and wait till
is displayed in the Unit Weight window. And then put
the corresponding weights on.