Zero Points is too low
Make sure the pan is on the scale or
perform the 3 points calibration.
Unstable Zero Point
(1)Make sure there is no winds or
vibration.
(2)Change the stability /vibration
parameter accordingly.
Battery for RTC is too weak
Replace a new battery or press
to
abort.
Overload (The weights on the pan
is more than +9e of the max.
capacity )
Take away the overloaded objects.
The value in the Total Weight
window or Total Count Window
exceeds the display range
(1)Press
key twice to clear the all
the accumulation or press
key to
return to normal weighing mode .
(2)Change another bigger unit weight
The last accumulation is more than
the preset accumulation allowed . Press
key twice to clear the all the
accumulation or press
key to return
to normal weighing mode .
8-13 Problem & Solution
Printer cannot print out
z
Make sure the parameter P-05 is set correctly.
z
Make sure the parameter P-06 and P-07 match the printer.
z
If EZ or BP is being used, they should have download the printer format before.
HI & LO checking do not work
z
Make sure the function is enabled.
z
The value of LO is greater than HI.
No sound in HI & LO checking function
Make sure the parameter P-03 is set correctly.
Reset the time every time when the scale is turned on
Connect an optional RTC
How to check whether the printer is compatible to the scale