4
100Ω RTD Temperature
Range
23 to 500°F (-5 to 260°C)
Resolution
0.1°F (0.1°C)
Accuracy
± 1% of Reading or ± 1°C, whichever is greater
1000Ω RTD Temperature
Range
23 to 500°F (-5 to 260°C)
Resolution
0.1°F (0.1°C)
Accuracy
± 1% of Reading or ± 0.3°C , whichever is greater
10k or 100k Thermistor Temperature
Range
23 to 194°F (-5 to 90°C)
Resolution
0.1°F (0.1°C)
Accuracy
± 1% of Reading or ± 0.3°C, whichever is greater
Analog (4-20 mA)
Range
0 to 22 mA
Resolution
0.01 mA
Accuracy ± 0.5% of reading
Electrodeless Conductivity
Range
Resolution
Accuracy
500-12,000 µS/cm
1 µS/cm, 0.01 mS/cm, 0.1 mS/m, 0.001 S/m, 1 ppm
1% of reading
3,000-40,000 µS/cm
1 µS/cm, 0.01 mS/cm, 0.1 mS/m, 0.001 S/m, 1 ppm
1% of reading
10,000-150,000 µS/cm
10 µS/cm, 0.1 mS/cm, 1 mS/m, 0.01 S/m, 10 ppm
1% of reading
50,000-500,000 µS/cm
10 µS/cm, 0.1 mS/cm, 1 mS/m, 0.01 S/m, 10 ppm
1% of reading
200,000-2,000,000 µS/cm
100 µS/cm, 0.1 mS/cm, 1 mS/m, 0.1 S/m, 100 ppm
1% of reading
Temperature °C
Range Multiplier
Temperature °C
Range Multiplier
0
181.3
80
43.5
10
139.9
90
39.2
15
124.2
100
35.7
20
111.1
110
32.8
25
100.0
120
30.4
30
90.6
130
28.5
35
82.5
140
26.9
40
75.5
150
25.5
50
64.3
160
24.4
60
55.6
170
23.6
70
48.9
180
22.9
Note: Conductivity ranges on page 2 apply at 25°C. At higher temperatures, the range is reduced per the range multiplier
chart.