15
O
per
ati
on
This section describes pump operation. Observe instructions in this manu-
al. See manufacturer's instruction manual for the motor driver.
Pump setting
First, program operation of the pump.
Pulse input direction & Motor rotation
The pump lets out liquid at the input of the CCW direction command pulse and takes in liquid at the input of the
CW direction command pulse.
The motor-driven cylinder is at the origin at factory default setting. Before operation, use a tester to check an
output of the home sensor is at the “L (GND)” level.
Number of pulses & Discharge capacity (RV)
Calculated flow rate
Number of input pulses
1ml
2400
2ml
4800
3ml
7200
4ml
9600
5ml
12000
*
The above data is based on the assumption that the driver is set to half step with the motor-driven cylinder at the origin.
Actual discharge capacity varies with a piping condition or so.
The lower limit of the motor-driven cylinder
NOTE
Do not move the motor-driven cylinder over the lower limit.
Operation
Pump setting
The lower limit of sensor dog
ON position
: Suction process (CW rotation)
Home sensor
Observe the lower limit, which comes to a clock-
wise rotation for 600 pulses from the ON position
Sensor dog
: Discharge process (CCW rotation)