5.4 Program structure
The Programat CS4 basically offers two types of
programs:
a) Standard program for materials of the
Ivoclar Vivadent group
(see enclosed
Program Table)
e.g. IPS e.max ZirCAD, IPS e.max CAD
b) Free programs
All the free programs are equal and available
as full-fledged programs. All the parameters
can be individually set in each program.
When the furnace is delivered ex
factory, the standard programs already
contain the recommended material
parameter settings.
The programs have been designed in such a
way that 2 predrying stages, 3 heating stages,
2 cooling stages, and 1 opening time can be set.
In case one of these stages is not
required, the time must be set to
00:00.
The predrying time includes the time of
the heating phase and is no holding
time.
5.5 Settable parameters and possible value ranges
* If the holding temperatures T2 and T3 are lower than 900 °C or 1652 °F , the maximum temperatures for the
temperature increase rates t2 and t3 are 130 °C/min or 234 °F/min.
Automatic plausibility check of the parameters
The Programat CS4 is equipped with an automatic plausibility check function.
The parameters are checked upon each program start.
In case of contradictory parameter combinations, the program stops automatically and
the respective error number is indicated.
Value range °C
Value range °F
Symbol Parameters
min.
max.
Unit
min.
max.
Unit
v1
Predrying stage 1
Predrying temperature
0/100
300
°C
0/212
572
°F
V1
Predrying time
00:00
00:20
hh:mm
00:00
00:20
hh:mm
v2
Predrying stage 2
Predrying temperature
0/100
300
°C
0/212
572
°F
V2
Predrying time
00:00
00:20
hh:mm
00:00
00:20
hh:mm
S
Closing time
Closing time
00:00
01:00
hh:mm
00:00
01:00
hh:mm
t1
Heating stage 1
Temperature increase
rate
10
130
°C/min
18
234
°F/min
T1
Holding temperature
300
900
°C
572
1652
°F
H1
Holding time
00:00
01:00
hh:mm
00:00
01:00
hh:mm
t2
Heating stage 2
Temperature increase
rate
10
50*
°C/min
18
90*
°F/min
T2
Holding temperature
0/300
1500
°C
0/572
2732
°F
H2
Holding time
00:00
01:00
hh:mm
00:00
01:00
hh:mm
t3
Heating stage 3
Temperature increase
rate
10
15*
°C/min
18
27*
°F/min
T3
Holding temperature
0/300
1600
°C
0/572
2912
°F
H3
Holding time
00:00
01:00
hh:mm
00:00
01:00
hh:mm
t4
Cooling stage 1
Temperature increase
rate
10
70
°C/min
18
126
°F/min
T4
Holding temperature
100
1200
°C
212
2192
°F
H4
Holding time
00:00
01:00
hh:mm
00:00
01:00
hh:mm
t5
Cooling stage 2
Temperature increase
rate
10
50
°C/min
18
90
°F/min
T5
Holding temperature
0/100
1100
°C
0/212
2012
°F
H5
Holding time
00:00
01:00
hh:mm
00:00
01:00
hh:mm
O
Opening time
Opening time
00:01
05:00
hh:mm
00:01
05:00
hh:mm
Graphic overview of the program parameters
Predrying
Heating
Cooling
Opening
V1
v1
v2
H1
H2
T2
T3
T4
T5
H3
H4
H5
T1
t1
t2
t3
t4
t5
V2