Parameters
Descriptions
Current
It is the current value set in this step when the mode is
USB PPS. The setting range is 0A~5.1A.
Fail HV
It is the judgment condition of highest voltage in this step.
If voltage is higher than this setting value, the test fails.
When OFF is checked, no judgment is made. Setting
range: 0V~21.000V
Fail LV
It is the judgment condition of lowest voltage in this step.
If voltage is lower than this setting value, the test fails.
When OFF is checked, no judgment is made. Setting
range: 0V~21.000V
Fail HC
It is the judgment condition of highest current in this step.
If current is higher than this setting value, the test fails.
When OFF is checked, no judgment is made. Setting
range: 0A~5.100A
Fail LC
It is the judgment condition of lowest current in this step. If
current is lower than this setting value, the test fails. When
OFF is checked, no judgment is made. Setting range:
0A~5.100A
Testing Time
It is the running time for this step.
Illustration
View the curve graph of this step.
E Load
Mode
Running mode of the electronic load: VOLTAGE, CUR-
RENT, POWER, RESISTANCE, SHORT, OCP
Current
It is the current setting value with setting range of
0~5.100A.
Voltage
It is the voltage setting value with setting range of
0~21.000V.
Power
It is the power setting value with setting range of
0~150.00W.
Resistance
It is the resistance setting value with setting range of
0~7,500.0Ω.
On Time
It is the loading time. When OFF is checked, the loading
time will not be calculated. The setting range is
0~1,000.00S. This setting time should be less than the
setting time in Testing Time.
Copyright © Itech Electronic Co., Ltd.
26