Second row: Display the voltage setting range and maximum current value
under the PPS mode.
Press
UP
and
DOWN
buttons simultaneously to exit existing PPS mode.
5. Press
START
to enter PPS mode, and display the existing voltage and cur-
rent setting values.
When the existing voltage value flicks, it represents that the voltage value is
in to-be-set state. Press
UP
or
DOWN
to adjust the output voltage value.
The adjustment step of voltage value is 0.02V.
•
Press
UP
or
DOWN
: The voltage increases/decreases by 0.02V each
time.
•
Press and hold
UP
or
DOWN
: The voltage increases/decreases by
0.02V per 0.1Sec.
6. Press
MODE
button to switch existing setting parameter to current. Press
this button again to switch the setting parameter between voltage and
current.
7. When the current flicks, it represents that the current value is in to-be-set
state. Press
UP
or
DOWN
to adjust the output current value. The adjustment
step of current value is 0.05A.
•
Press
UP
or
DOWN
: The current increases/decreases by 0.05A each
time.
•
Press and hold
UP
or
DOWN
: The current increases/decreases by 0.05A
per 0.1Sec.
8. Press
START
button to change the output voltage and current based on the
setting voltage and current.
9. Turn on the
On/Off
button of load input control.
2.2.4 Test under List Mode (only for IT-E255A)
Only IT-E255A has test function under List mode. The List file can only be edited
in the host computer software. The user needs to use the host computer soft-
ware to edit the List file, and save the edited List file into the quick charger
Copyright © Itech Electronic Co., Ltd.