AC Voltage, ACV+DCV[3] (AC Coupled)
Range
Resolu
-tion
Full
Scale
Accuracy (1 year 23
˚
C ±5
˚
C) [1]
30-50Hz 50-10kHz 10K-30kHz 30K-100kHz
500mV 10
μ
V 510.00 1.00%+40 0.50%+40 2.00%+60 3.00%+120
5V 100
μ
V 5.1000 1.00%+20 0.35%+15 1.00%+20 3.00%+50
50V 1mV
51.000
1.00%+20 0.35%+15 1.00%+20 3.00%+50
500V 10mV 510.00
x
0.5%+15 1.00%+20[2]
3.00%+50[2]
750V 100mV 765.0
x
0.5%+15
x
x
[1]Specifications are for sine wave inputs that are greater than 5% range.
[2]Input voltage <300Vrms.
[3]The accuracy of ACV+DCV is equal to ACV’s with 10 more digits added.
* The specifications are guaranteed to an input of 750V. A beeping alarm
will go off when the input value is higher than 750V.
* Input protection of 1000V peak on all ranges.
* AC-coupled true RMS – measures the AC component of the input with up
to 400Vdc of bias on any range.
* AC Common Mode Rejection Ratio.
>60 dB at dc, 50 or 60Hz ± 0.1% (1k
Ω
unbalanced, slow rates)
*Input impedance 1M
Ω
±2% in parallel with 100pF.
AC Current, ACI+DCI[3] (AC Coupled)
Range
Resolu
-tion
Full
Scale
Accuracy (1 year 23
˚
C ±5
˚
C) [1]
Burden
Voltage
30-50Hz
50-2kHz
2K-5kHz
5K-20kHz
500
μ
A 10nA 510.00 1.50%+50 0.50%+40 1.50%+50 3.00%+75 0.06V
max
5mA 100nA 5.1000 1.50%+40 0.50%+20 1.50%+40 3.00%+60 0.6V
max
50mA 1
μ
A 51.000 1.50%+40 0.50%+20 1.50%+40 3.00%+60 0.14V
max
500mA 10
μ
A 510.00 1.50%+40 0.50%+20 1.50%+40 3.00%+60[2] 1.4V max
5A 100
μ
A 5.1000 2.0%+40 0.50%+30
x
x
0.5V max
10A 1mA
12.000 2.0%+40 0.50%+30
x
x
0.8V
max
[1] The 500
μ
A range requires an input of >35
μ
A to meet specifications. The
5mA~10A ranges need more than 5% of full scale range to meet
specifications.
[2] Input current (5k ~ 20kHz)<330mArms.
[3]The accuracy of ACI+DCI is equal to ACI’s with 10 more digits added.
Summary of Contents for IDM-8341 Series
Page 1: ...Quick Start Guide Dual Measurement Multimeter IDM 8341 8342 Series EN...
Page 10: ...DC Voltage Range 0 to 1000V Connection Activation Auto range setting Manual range setting...
Page 11: ...AC DC Voltage Range 0 to 1000V Connection Activation Auto range setting Manual range setting...
Page 13: ...DC Current Range 0 to 2A Connection Activation Auto range setting Manual range setting...
Page 14: ...AC DC Current Range 0 to 2A Connection Activation Auto range setting Manual range setting...
Page 18: ...Temperature Range Thermocouple 200 C 300 C Connection Activation...
Page 21: ...Max Applicable in Activation Min Applicable in Activation x 2...
Page 24: ...4 Confirm low setting...