* The specifications are guaranteed to 10A. A beeping alarm will go off when
the input current being measured is higher than 10A.
Resistance
Resistance Resolution Full Scale
Test Current
Accuracy (1 year
23
˚
C ±5
˚
C)[2]
500
Ω
10m
Ω
510.00 0.83mA 0.1%+5
[1]
5k
Ω
100m
Ω
5.1000 0.83mA
0.1%+3
[1]
50k
Ω
1
Ω
51.000 83
μ
A 0.1%+3
500k
Ω
10
Ω
510.00 8.3
μ
A 0.1%+3
5M
Ω
100
Ω
5.1000 830nA
0.1%+3
50M
Ω
1K
Ω
51.000
560nA//10M
Ω
0.3%+3
[1] Using the REL function. If you don't use the REL function then increase
the error by 0.2
Ω
.
[2] When measuring resistances greater than 500k
Ω
, please use shielded
test leads to eliminate the noise interference that may be induced by
standard test leads.
* Open circuit voltage approximates 6V max on 500~5M
Ω
range,
approximates 5.5V max on 50M
Ω
range.
* Input protection of 500V peak on all ranges.
Diode
Range
Resolution
Full Scale
Test Current
Accuracy(1 year
23
˚
C ±5
˚
C)
5V 100
μ
V 5.1000 0.83mA
0.05%+5
* Input protection of 500V peak. *Open circuit voltage approximates 6V.
Continuity
Range
Resolution Full Scale
Test Current
Accuracy(1 year
23
˚
C ±5
˚
C)
5000.0
Ω
100m
Ω
5100.0 0.83mA
0.1%+5
* Input protection of 500V peak. *Open circuit voltage approximates 6V.
Summary of Contents for IDM-8341 Series
Page 1: ...Quick Start Guide Dual Measurement Multimeter IDM 8341 8342 Series EN...
Page 10: ...DC Voltage Range 0 to 1000V Connection Activation Auto range setting Manual range setting...
Page 11: ...AC DC Voltage Range 0 to 1000V Connection Activation Auto range setting Manual range setting...
Page 13: ...DC Current Range 0 to 2A Connection Activation Auto range setting Manual range setting...
Page 14: ...AC DC Current Range 0 to 2A Connection Activation Auto range setting Manual range setting...
Page 18: ...Temperature Range Thermocouple 200 C 300 C Connection Activation...
Page 21: ...Max Applicable in Activation Min Applicable in Activation x 2...
Page 24: ...4 Confirm low setting...