DC Voltage
Range Resolution Full
Scale
Accuracy (1 year
23
˚
C ±5
˚
C) Input
Resistance
500mV 10
μ
V 510.00
0.02%+4
10M
Ω
or >10G
Ω
5V 100
μ
V 5.1000
10M
Ω
or >10G
Ω
50V 1mV 51.000
11.1M
Ω
500V 10mV 510.00
10.1M
Ω
1000V 100mV 1020.0
10M
Ω
* When the input value exceeds the full scale of the selected range, the
display will show -OL- (over load) on the display.
* The specifications are guaranteed to an input voltage of 1000V. A beeping
alarm will go off when the input voltage is higher than 1000V.
* Input protection of 1000V peak on all ranges.
* DC Common Mode Rejection Ratio
>90 dB at dc, 50 or 60Hz ± 0.1% (1k
Ω
unbalanced, slow rates)
DC Current
Range Resolution Full Scale
Accuracy
(1 year 23
˚
C ±5
˚
C)
Shunt
Resistance
Burden
Voltage
500
μ
A 10nA 510.00
0.05%+5
100
Ω
0.06V
max
5mA 100nA 5.1000
0.05%+4
100
Ω
0.6V
max
50mA 1
μ
A 51.000 0.05%+4
1
Ω
0.14V
max
500mA 10
μ
A 510.00 0.10%+4
1
Ω
1.4V
max
5 A
100
μ
A 5.1000
0.25%+5
10m
Ω
0.5V
max
10 A
1mA
12.000
0.25%+5
10m
Ω
0.8V
max
* 500
μ
A~500mA range has a 3.6V voltage limit protection and 0.5A fuse
protection. And 10A range has a 12A fuse protection.
* When the input value exceeds the full scale of the selected range, the
display will show -OL- (over load) on the display.
* The specifications are guaranteed to an input of 10A. A beeping alarm will
go off when the input value is higher than 10A.
Summary of Contents for IDM-8341 Series
Page 1: ...Quick Start Guide Dual Measurement Multimeter IDM 8341 8342 Series EN...
Page 10: ...DC Voltage Range 0 to 1000V Connection Activation Auto range setting Manual range setting...
Page 11: ...AC DC Voltage Range 0 to 1000V Connection Activation Auto range setting Manual range setting...
Page 13: ...DC Current Range 0 to 2A Connection Activation Auto range setting Manual range setting...
Page 14: ...AC DC Current Range 0 to 2A Connection Activation Auto range setting Manual range setting...
Page 18: ...Temperature Range Thermocouple 200 C 300 C Connection Activation...
Page 21: ...Max Applicable in Activation Min Applicable in Activation x 2...
Page 24: ...4 Confirm low setting...