S
PECIFICATIONS
The specifications apply when the DMM is warmed up for at least 30
minutes and operates in slow rate.
Below are the basic conditions required to operate the DMM within
specifications:
•
Calibration:
Yearly
•
Operating Temperature Specification: 18~28
˚
C (64.4~82.4
˚
F)
•
Relative Humidity: 80% (Non condensing)
•
Accuracy: ± (% of R Digits)
•
AC measurements are based on a 50% duty cycle.
•
The power supply cable must be grounded to ensure accuracy.
•
All specifications are applicable to the main (1
st
) display only.
General Specifications
Specification Conditions:
Temperature: 23
˚
C±5
˚
C
Humidity: <80%RH, 75%RH for resistance measurement readings greater
than 10M
Ω
.
Operating Environment: (0~50
˚
C)
Temperature Range: 0~35
˚
C, Relative Humidity: <80%RH;
>35
˚
C, Relative Humidity: <70%RH
Indoor use only
Altitude: 2000 meters
Pollution degree 2
Storage Conditions (-10~70
˚
C)
Temperature Range: 0~35
˚
C, Relative Humidity: <90%RH;
>35
˚
C, Relative Humidity: <80%RH
General:
Power Consumption: Max 15VA
Dimensions: 265 mm (W) X 107 mm (H) X 302 mm (D)
Weight: Approximately 2.9 kg
Summary of Contents for IDM-8341 Series
Page 1: ...Quick Start Guide Dual Measurement Multimeter IDM 8341 8342 Series EN...
Page 10: ...DC Voltage Range 0 to 1000V Connection Activation Auto range setting Manual range setting...
Page 11: ...AC DC Voltage Range 0 to 1000V Connection Activation Auto range setting Manual range setting...
Page 13: ...DC Current Range 0 to 2A Connection Activation Auto range setting Manual range setting...
Page 14: ...AC DC Current Range 0 to 2A Connection Activation Auto range setting Manual range setting...
Page 18: ...Temperature Range Thermocouple 200 C 300 C Connection Activation...
Page 21: ...Max Applicable in Activation Min Applicable in Activation x 2...
Page 24: ...4 Confirm low setting...