21
Mode
Type
Action
Linear
0% to 100% to 0% per 40 sec
Linear
0% to 100% to 0% per 20 sec
25% Step
0% to 100% to 0%, a step per 15 sec
25% Step
0% to 100% to 0%, a step per 5 sec
Input Protection :
Equipped with High Energy Fuse.
440mA, 1000V IR 10kA Fuse (Bussmann DMM-B-44/100)
Power Source :
Source Mode : Internal batteries, > 4.5V
Simulate Mode : External loop supply, 6V to 48V
Output Short Protection
11. Loop Power
Function
Range
Driver
Accuracy
Normal
50.000mA
30V / 1.25k
Ω
± (0.05% + 5d)
250
Ω
Hart
50.000mA
24V / 1k
Ω
± (0.05% + 5d)
Input Protection :
Equipped with High Energy Fuse.
440mA, 1000V IR 10kA Fuse (Bussmann DMM-B-44/100)
Power Source :
Internal batteries, > 4.5V
Minimum Output Voltage :
24V
Output Short Protection
9. Auto DC Current Output
Range
Resolution
Accuracy
0.000mA to 20.000mA
Over range to 24.000mA
± (0.5% + 5d)
4.000mA to 20.000mA
Over range to 24.000mA
1uA
Input Protection :
Equipped with High Energy Fuse.
440mA, 1000V IR 10kA Fuse (Bussmann DMM-B-44/100)
Power Source :
Source Mode : Internal batteries, > 4.5V
Simulate Mode : External loop supply, 6V to 48V
Output Short Protection
10. Auto DC Current Output