background image

1.  About  

ITA/ITB-AS/AP/AW Series Table Top Scales

We would like to thank you for choosing us as your preferred choice for your weighing 
requirements.

The ITA/ITB -AS/AP/AW series weighing scale is an accurate, fast and versatile series of 
splash proof / water proof Table Top scales, loaded with time saving features and ease 
of customer use. 

The Scale is loaded with features like, automatic zero tracking, overload alarm, low 
battery protection, pre-set tare and an accumulation function to add pieces being 
counted and finally recall the accumulated total.

The  Scale  features  auto  zero  tracking  to  automatically  compensate  for  small 
fluctuations in zero such as a build up of material on the platform

To extend battery life an auto-sleep feature is standard the operator can set the auto-
sleep  ON  to  switch  of  the  scale  display  when  idle.  Alternatively  the  auto-sleep 
function  can  be  switched  OFF,  leaving  the  scale  display  permanently  ON  until 
switched off by the operator.

ITA Series – Table Top Precision Weighing Balance

ITA Series – Table Top Precision Weighing Balance

2. TECHNICAL DATA

12

1

Basic Parameters

None

Measurement Data

Capacity

Selectable 3kg ~ 30 kg

Readout / Interval

100mg ~ 5g based on capacity

Tare range subtractive

100% subtractive

Repeatability

+- 1d

Max Display Resolution

3,000 ~ 100,000 Divisions

Application Modes

Weighing, Counting, Hi Lo OK alarm setting

Product Features

Pre-Set Tare, Piece Counting, Glow Control

Keyboard

4 membrane keys (On/off, Mode, Tare and Zero)

Display Indicator 

Stability, Zero Position, Tare, Charging Status, 
Accumulation, Percentage mode,
 function active, d=e/10 mode

Weighing Units

Kg

Calibration 

Digital with external weight.

Weight

4 kg Approx

Pan Size

230mm x197mm

Front Display Type

0.8" Bright red LED display 

Rear Display

Optional, Available with 0.56" LED Display

Communication
Stabilization

3 seconds

Power

9V AC Adaptor or Power Cord / built in battery for 
backup to 30hrs. / 40 hrs with sleep mode active

Operating Temp.

50° to 98° F / 10° to 35° C

Dimensions 

245 x 210 x 140 (w x d x h)

SOLUTION

ERROR DISPLAY

ERROR

8. ERROR CODES

·

 

Zero not in range

·

 

Weight not enough for 

the selected setting

·

 

LOW  Battery.  Machine 

displays  message  and 
turns off

Remove  weight  from  pan 
and restart the machine. If 
the  problem  persists,  the 
machine  may  have  been 
overloaded.

Keep  weight  or  Add  more 
weight  on  pan  for  correct 
results.

Battery  is  low.  Connect 
scale to power so that the 
battery can be charged.

 

 

 

REMEDY

SYMPTOM

POSSIBLE CAUSE

9. TROUBLE SHOOTING

Scale will not 
power on

Battery does not 
give enough
 backup

Weight reading on 
scale does not 
stabilize

Scale does not 
show weight 
accurately

External calibration 
is not stored after 
switch on / off the 
machine

If  running  on  batteries,  the  battery 
may have fully discharged. AC Power 
cord not connected. 

The  battery  is  defective.  The  battery 
may have past its useful life. 

Improper calibration.

This  happens  when  the  Internal  to 
External  calibration  variance  larger 
than 2%. The cause could be that the 
user is not using correct weight.

External  Rubbing  /  Friction.  A  scale 
cannot measure accurately if an object 
is  rubbing  or  pressing  against  the 
scale  platform.  Mobile  interference, 
Vibration  or  wind  could  be  other 
possible reasons.

Connect  scale  to 
p o w e r   s o u r c e .  
Recharge battery.

Change the battery.

Move the scale away 
from  mobile  signal, 
breeze  or  vibration 
areas. 

Calibrate  the  scale 
again  to  and  check 
for  the  problem 
again.

Use  correct  weights 
to  calibrate.  If  the 
problem  persists, 
calibrate  the  scale 
u s i n g   i n t e r n a l  

 

 

 

Summary of Contents for ITA SERIES

Page 1: ...Guide ISHTAA SCALES INC INDIA www ishtaascales com ITA ITB AS AP AP SERIES Electronic Table Top Weighing Balance ITA Series Table Top Precision Weighing Balance ITA Series Table Top Precision Weighing Balance TM 14 ...

Page 2: ... and purposeful destruction are NOT covered by this warranty Technical Service MUST be performed by authorised staff or factory authorized personnel Manufacturer s obligation is limited to replacement of parts that have been so returned and are disclosed to Manufacturer s satisfaction to be defective The provisions of these warranty clauses are lieu of all other warranties expressed or implied and...

Page 3: ...ernal weight Weight 4 kg Approx Pan Size 230mm x197mm Front Display Type 0 8 Bright red LED display Rear Display Optional Available with 0 56 LED Display Communication Stabilization 3 seconds Power 9V AC Adaptor or Power Cord built in battery for backup to 30hrs 40 hrs with sleep mode active Operating Temp 50 to 98 F 10 to 35 C Dimensions 245 x 210 x 140 w x d x h SOLUTION ERROR DISPLAY ERROR 8 ER...

Page 4: ...rectly when the bubble scale is in the center of the circle The unit comes with 12V AC adaptor or 220V AC as per model specifications country of use The scale is fitted with a built in 6V 4Ah battery which provides backup of more than 35 50 hours depending on glow setting and sleep mode setting Put a known weight on the machine and make sure that the scale reads the correct weight on the display I...

Page 5: ...ATION Internal calibration also referred as service calibration requires opening the machine and changing the jumpers It is advisable to do this step only if you have required assistance or the service personnel assisting you to do this Open the battery cover from the bottom of the machine and remove the cover and change the calibration jumper marked as CAL to ON Make sure there is nothing kept on...

Page 6: ...e changes geographic gravity variations altitude changes and abuse are few reasons why a scale may need recalibration The scale can be calibrated using a known external weight 4 9 7 1 EXTERNAL USER CALIBRATION To start calibration turn the scale off and then turn it on During the count down press and hold down the Zero button do not release until the display shows CAL Make sure there is no weight ...

Page 7: ...ounting function to come in to the weighing mode It is important to know that SLEEP mode increases the battery back up time by double The Normal backup time when the sleep mode is switched off is 30 Hours and with the SLEEP mode switched on the battery back up time increases to more than 60 Hrs If the USER wants to switch off the SLEEP MODE for some special usage or function it can be done as foll...

Page 8: ...s 3d ITA Series Table Top Precision Weighing Balance ITA Series Table Top Precision Weighing Balance 6 7 6 3 LED GLOW The display GLOW strength can be increased or decreased depending upon the user requirement Increasing the GLOW of the display will effect the Battery back up time and the battery life The GLOW setting is set to minimum by Default If the user wants to increase the GLOW of the displ...

Reviews: