5.  USER OPERATION

5.1 SWITCHING SCALE - ON / OFF

The scale is switched ON / OFF by 

pressing  the  Power  button:  This 

button  is  used  to  switch  on  and 

switch off the scale. It is a real ON/OF which is implemented using the latest 

technology so that there is absolutely no power consumption when the 

scale is switched off. 

5.2

MANUAL TARE

5.2.1

TARE button: This button is used 

for  taring  a  weight  to  display 

0.000. A Tare value is indicated 

by  the  “t”  LED  on  the  panel. 

Place  a  container  on  the 

weighing  pan  (say  200g),  then 

press  the  TARE  button.  The 

weight is tared and the scale will 

display zero and the tare LED will 

be  ON.  To  recall  the  tared 

weight,  simply  press  the  TARE 

button again. To TARE the value, 

again press the TARE .

ITA Series – Table Top Precision Weighing Balance

ITA Series – Table Top Precision Weighing Balance

7 CALIBRATION 

All  weighing  scales  come  pre-calibrated  from  the  factory.  However  the 

scale may require re-calibration. Temperature changes, geographic gravity 

variations, altitude changes and abuse are few reasons why a scale may 

need  recalibration.  The  scale  can  be  calibrated  using  a  known  external 

weight.

4

9

 

 

 

 

 

 

7.1

EXTERNAL USER CALIBRATION 

To start calibration turn the scale off and 
then turn it on. During the count down, 
press and hold down the Zero button (do 
not release until the display shows “CAL”.

Make sure there is no weight on the pan. 
Press  Zero  key  and  the  display  shows 
“noLoAd”  momentarily  and  shows  the 
counts at zero level.
 
Pause  for  2  seconds  at  the  previous  step 
and  press  Zero  key.  The  display  shows 
“LoAd”  momentarily  and  shows  the 
internal counts. 

Put a known weight on the weighing pan 
(Ex.: 10kg). Press the ZERO key. The display 
will now show all “000.000” with the units 
digit blinking. 

If you put 10kg weight, Enter the weight in 
grams “010000”. Change the blinking digit 
by  pressing  the  tare  button.  Shift  the 
blinking digit to left by pressing the mode 
button once.

Press the Zero key once again. Display will 
show “END” 

With the weight still on the pan, press Zero 
key  once  again  to  return  to  weighing 
mode. Check calibration.

STEP

DISPLAY

PROCEDURE

1.

2.

3.

4.

5.

6.

7.

 

 

 

 

 

 

 

 

 

Summary of Contents for ITA SERIES

Page 1: ...Guide ISHTAA SCALES INC INDIA www ishtaascales com ITA ITB AS AP AP SERIES Electronic Table Top Weighing Balance ITA Series Table Top Precision Weighing Balance ITA Series Table Top Precision Weighing Balance TM 14 ...

Page 2: ... and purposeful destruction are NOT covered by this warranty Technical Service MUST be performed by authorised staff or factory authorized personnel Manufacturer s obligation is limited to replacement of parts that have been so returned and are disclosed to Manufacturer s satisfaction to be defective The provisions of these warranty clauses are lieu of all other warranties expressed or implied and...

Page 3: ...ernal weight Weight 4 kg Approx Pan Size 230mm x197mm Front Display Type 0 8 Bright red LED display Rear Display Optional Available with 0 56 LED Display Communication Stabilization 3 seconds Power 9V AC Adaptor or Power Cord built in battery for backup to 30hrs 40 hrs with sleep mode active Operating Temp 50 to 98 F 10 to 35 C Dimensions 245 x 210 x 140 w x d x h SOLUTION ERROR DISPLAY ERROR 8 ER...

Page 4: ...rectly when the bubble scale is in the center of the circle The unit comes with 12V AC adaptor or 220V AC as per model specifications country of use The scale is fitted with a built in 6V 4Ah battery which provides backup of more than 35 50 hours depending on glow setting and sleep mode setting Put a known weight on the machine and make sure that the scale reads the correct weight on the display I...

Page 5: ...ATION Internal calibration also referred as service calibration requires opening the machine and changing the jumpers It is advisable to do this step only if you have required assistance or the service personnel assisting you to do this Open the battery cover from the bottom of the machine and remove the cover and change the calibration jumper marked as CAL to ON Make sure there is nothing kept on...

Page 6: ...e changes geographic gravity variations altitude changes and abuse are few reasons why a scale may need recalibration The scale can be calibrated using a known external weight 4 9 7 1 EXTERNAL USER CALIBRATION To start calibration turn the scale off and then turn it on During the count down press and hold down the Zero button do not release until the display shows CAL Make sure there is no weight ...

Page 7: ...ounting function to come in to the weighing mode It is important to know that SLEEP mode increases the battery back up time by double The Normal backup time when the sleep mode is switched off is 30 Hours and with the SLEEP mode switched on the battery back up time increases to more than 60 Hrs If the USER wants to switch off the SLEEP MODE for some special usage or function it can be done as foll...

Page 8: ...s 3d ITA Series Table Top Precision Weighing Balance ITA Series Table Top Precision Weighing Balance 6 7 6 3 LED GLOW The display GLOW strength can be increased or decreased depending upon the user requirement Increasing the GLOW of the display will effect the Battery back up time and the battery life The GLOW setting is set to minimum by Default If the user wants to increase the GLOW of the displ...

Reviews: