RCS
Read Current Setpoint
– Reads the setpoint for the LD
current. The response is the command echoed back, followed
by a delimiter of “: “, and then the current setpoint in %.
Sent:
“RCS”
Response:
“RCS: 56.7”
(Indicates that the LD current
setpoint is 56.7%)
RCT
Read Laser Temperature
– Reads the internal temperature of
the laser. The response will be an echo of the command, a
delimiter of “: “, and the temperature in degrees centigrade.
Sent:
“RCT”
Response:
“RCT: 34.5”
RERR
Reset Errors
– Resets any resettable errors.
Sent:
“RERR”
Response:
“RERR”
RET
Read Elapsed Time
– Reads the elapsed time the laser has
been ON. The time is returned in minutes.
Sent:
“RET”
Response:
“RET: 1105”
RFV
Read current software revision.
Sent:
“RFV”
Response:
“RFV: 1.0.147 “
RIP
Read IP
– Reads the current IP address of the device. The
response is the command echoed back, followed by a
delimiter of “: “, and then the IP address in dot-decimal
notation.
Sent:
“RIP”
Response:
“RIP: 10.0.0.2”
RMASK
Read Subnet Mask
– Reads the current subnet mask of the
device. The response is the command echoed back, followed
by a delimiter of “: “, and then the subnet mask in dot-decimal
notation.
Sent:
“RMASK”
Response:
“RMASK:
255.0.0.0”
RMEC
Read Module Error Code. Returns error code stored in the
laser or zero if normal operation.
Sent:
“RMEC”
Response:
“RMEC: 0”
RNC
Read Minimum Current Setpoint
– Reads the minimum
current setpoint that can be set in the laser. The response will
be the command echoed back, followed by a delimiter of “: “,
then the minimum current as a percentage of the maximum.
Sent:
“RNC”
Response:
“RNC: 10.0”
(Indicates that the minimum
setpoint is 10.0 %)