English
- 20 -
Remedy
Eliminate all extension hoses and connect directly to spigot with
a single hose.
Check for blockage or kinks in hose.
Connect to different water source.
Clean or replace sediment filter.
Replace carbon filter.
Contact the service centre.
Problem
High wastewater flow with low pure water flow good system
pressure.
Remedy
Remove DI resin cartridge and retest.
Extremely cold water supply.
Plugged membrane - Replace membrane.
Problem
High pure water TDS
Remedy
Damaged membrane - replace membrane.
Membrane installed upside down.
Problem
Low System pressure with low flush water flow and/or low
waste water flow.
Remedy
Verify supply water conditions.
Clean or replace sediment filter
Replace carbon block filter.
Check pump relief valve setting (Electric unit only).
SCRAPPING
If you decide to stop using the equipment, you should render it
inoperative by removing the electric power cable. Parts which
might constitute a danger, especially to children playing with
the appliance, should be made harmless.
The equipment is classified as special waste. It must be taken
apart and divided into uniform sections which should be dis
-
posed off according to current legislation.
Do not use the components removed as spare parts.
BATTERIES
Considering the danger of some components, it is advisable
to take them to an authorized collection and disposal center.
TROUBLESHOOTING
Problem
Electric motor does not start.
Remedy
Check outlet to ensure it has live power.
Remove extension cord(s) if being used. Plug the machine
directly into outlet.
Test and reset the machine.
Unplug the machine and check cord wire connections.
Inspect cord for damage.
Contact the service centre.
Problem
Low supply water flow or pressure.
TECHNICAL CHARACTERISTICS
Without Pump Unit
With Electric Motor Unit
With Battery Unit
Dimensions
50 cm (20")x53 cm
(21")x127 cm (50")
50 cm (20")x53 cm
(21")x127 cm (50")
50 cm (20")x53 cm
(21")x127 cm (50")
Weight (Without Filters)
23 kg, (50 lbs)
53 kg, (117 lbs)
53 kg, (117 lbs)
Power supply.
-
1ac 230V 50Hz (60Hz)
cc 12V 33Ah
Inlet Pressure.
3 - 6 bar (44 - 90 psi)
1,5 - 6 bar (22 - 90 psi)
1,5 - 6 bar (22 - 90 psi)
Water Pressure.
Water Pressure supply
About 6 - 7 bar (90 - 100 psi)
About 4 - 5 bar (60 - 75 psi)
Pure Water production.
1,5 - 2,5 lpm ( 0,40 - 0,7
gpm) Medium
5 lpm (1,32 gpm) med.
2 - 3,5 lpm (0,53 - 0,93 gpm)
med.
Waste water production.
2 lpm (0,53 gpm) abt.
3 lpm (0,80 gpm) abt.
2 lpm (0,53 gpm) abt.
Stream of water maximun
height.
Depending on tap water
pressure maximum about 10
m (33 ft) with one pole.
20 m (65 ft) with 2 poles
20 m (65 ft) with 1 pole
Noise
Lp dB (A) 63 (kpA. 2dB)
Lw dB (A) 72 (kwA. 2dB)
Lp dB (A) 62 (kpA. 2dB)
Lw dB (A) 71 (kwA. 2dB)
Pump Unit System.
Motor RPM 550W a
1725/1425 rpm, 50/60 hz,
115/110V - 208V - 220/230V,
Pump
type
PRG8AH
800LT/H S/BY-P OT max 16
bar.
Battery 12 V 74 Ah C5
motor type ME71036 12V
IP65 150W
Pump 300LT/H S/BY-P OT.
Battery Charger: 12 V 10 A.
Summary of Contents for HighPure HP B
Page 54: ...54 IEC 60364 1 1 5 IEC 60364 1 I...
Page 55: ...55 30 30...
Page 58: ...58 E IEC 60364 1 1 2 3 4 2 5 6 fig 6 1 2 3 I 4 5 10 6 7 5 7 1 2 3 4 1 2 0 3 4 5...
Page 62: ...62...