Gasclam
Ion Science Ltd
Page 35 of 37
Technical Specification
Sensor
Method/type
Range
Resolution
Accuracy
Linearity
CH
4
Infrared
0-100 %
1% of measuring
range above
50%, 0.5% below
50%.
± 2% FSD
+/- 2% FSD or 10% reading
CO
2
Infrared
0-100 %
1% of measuring
range above
50%, 0.5% below
50%.
± 2% FSD
+/- 2% FSD or 10% reading
Oxygen
Electrochemical
0-25 %
0.1 %
± 5% of reading ± 1
digit
>1 % O
2
deviations @ 10%
O
2
CO*
Electrochemical
0-1000 PPM
1 PPM
<± 6ppm at 0, ± 5%
at 250 ppm
Linear at 0 and 400ppm,
error at full scale <40ppm
H
2
S*
Electrochemical
0 - 100
1 PPM
<± 1 ppm at 0, ±
5% at 50 ppm
Linear at 0 and 20ppm,
error at full scale 0 to -9ppm
VOC*
PID
0-4,000 PPM
1 PPM
± 5% of reading ± 1
digit
+/- 5% to 100 ppm
Other range and resolution CH
4
and CO
2
sensors are available on request.
Environmental
Method / Type
Range
Resolution
Barometric Pressure
Piezoelectric
800-1200 mBar
1 mBar
Borehole Pressure
Piezoelectric
800-1200 mBar
1 mBar
Temperature
Internal Chip
-5°C to +50°C or 41°F to 122°F
1°C or 1°F
Water depth*
Piezoelectric
0 – 25 m
0.01m
* Optional
Memory
65000 time / date stamped readings
Power
Internal x 2 Alkaline D-cells or rechargeable battery pack
Battery Life
Approximately 1 month (Based on hourly sampling with rechargeable battery pack)
Case
High Quality Stainless Steel
Weight
6 kg or 13.2 lb
Protection
IP – 68 (continuous submersion)
Operation
–5 - +50 °C or 41°F to 122°F
Approvals
CE, EMC, ATEX, 0105 X, Ex II 2G, Ex d ib [ib] IIB T4
Certification rating
Ex 2G Ex d ib [ib] IIb T4
Certificate number
FTZU 07 ATEX 0105 X
Specifications obtained under laboratory conditions of gas flow, temperature and humidity. Field performance is
dependent on the correct installation procedures been followed, environmental conditions, frequent sensor cleaning and
regular calibration. Some of the sensors have cross sensitivity, for more information contact Ionscience.
Summary of Contents for GasClam
Page 1: ...GasClam Instrument User Manual V1 5...
Page 2: ...Gasclam Ion Science Ltd Page 1 of 37...
Page 10: ...Gasclam Ion Science Ltd Page 9 of 37 Getting Started How the GasClam Works and access areas...
Page 26: ...Gasclam Ion Science Ltd Page 25 of 37 Software Displayed multiple parameters...
Page 31: ...Gasclam Ion Science Ltd Page 30 of 37 Other tests This section is for service centres...
Page 35: ...Gasclam Ion Science Ltd Page 34 of 37 Service and Calibration Spare parts for users...