Gasclam
Ion Science Ltd
Page 6 of 37
The effects of water on your GasClam
2. Effects of flooding on data and GasClam
When the unit is not sampling it has an ingress protection rating of IP-68 (submersible to 10 meters for 1 week).
However,
Immersion will effect data and
May require subsequent corrective action
These effects and requirements will differ whether immersed from above, i.e. flooded headworks, or from below, i.e. rise
in borehole water level, see table below:
Flooded
Headworks
Remedial action
Preventive maintenance
Effects on
data
should vent breather be
submerged:
atmospheric pressure
reading will be incorrect
and will remain incorrect
Scheduled venting will be
ineffective.
Use borehole pressure as approximate
replacement or use data from another
source.
If chamber is known to
flood, or is already full of
water
Chose another
chamber
Or
replace
snorkel with
bolt
Effects on
GasClam
If submerged for an extended
period the vent pipe may become
flooded.
The vent breather may become
clogged by dirty water.
When GasClam is visited
Check data and call service
centre if atmospheric
pressure looks strange and
they will advise
If chamber is liable to
flooding
replace vent
breather each visit
Flooded
Borehole
Remedial action
Preventive maintenance
Effects on
data
Should water reach the bottom of
the GasClam the inflow and
outflow valves will remain closed
and no samples will be taken:
borehole pressure reading
will be incorrect
gas concentration
readings will be incorrect
Data points collected whilst the
sampling cycle was disabled will be
marked
If borehole water level is
known to be high or is
already full of water
Chose another
chamber
Or extend the
borehole above
ground level
Effects on
GasClam
None
None
Summary of Contents for GasClam
Page 1: ...GasClam Instrument User Manual V1 5...
Page 2: ...Gasclam Ion Science Ltd Page 1 of 37...
Page 10: ...Gasclam Ion Science Ltd Page 9 of 37 Getting Started How the GasClam Works and access areas...
Page 26: ...Gasclam Ion Science Ltd Page 25 of 37 Software Displayed multiple parameters...
Page 31: ...Gasclam Ion Science Ltd Page 30 of 37 Other tests This section is for service centres...
Page 35: ...Gasclam Ion Science Ltd Page 34 of 37 Service and Calibration Spare parts for users...