![Invitrogen E-Gel Power Snap Plus User Manual Download Page 70](http://html1.mh-extra.com/html/invitrogen/e-gel-power-snap-plus/e-gel-power-snap-plus_user-manual_2085413070.webp)
Related products
Accessory products
Choosing the right gel
To obtain the best results for your application, it is important to choose the correct agarose percentage
and well format. Go to
thermofisher.com/Egel-ladderselectiontool
for additional information, or use
the specifications of various single-use only gels in the followng tables to determine the optimal gel
type for your purpose.
Gels for preparative gel electrophoresis in cloning and NGS applications
E
‑
Gel
™
CloneWell
™
II
E
‑
Gel
™
Size Select II
E
‑
Gel
™
NGS
Application
Target fragment isolation
in cloning workflow
Low range fragment
library size selection in
NGS workflow
High range fragment
library size selection
No rows
2 rows: 1 loading row and
1 recovery row
2 rows: 1 loading row and
1 recovery row
1 row with sample
loading wells
Loading wells
7
7
10 + 1 marker lane
Loading volume
25 µL
25 µL
20 µL
Stain
SYBR
™
Safe
SYBR
™
Gold II
SYBR
™
Safe
Detection sensitivity
5 ng/band
0.5 ng/band
5 ng/band
% Agarose
0.8%
2%
0.8%
Separation range
100 bp – 6 kb
50 bp – 2 kb
800 bp – 10 kb
Run time
13-28 min
8-20 min
26-32 min
Access to sample
Sample recovered via
elution wells
Sample recovered via
elution wells
Openable cassette.
Manual gel excision.
G
70
E
‑
Gel
™
Power Snap Plus Electrophoresis System User Guide