8.2 Muting the Alarms
Press the Mute button to mute alerts and alarms.
Pressing the mute button will only silence the alarm for 24 hours if the alarm condition
remains active during that time. If the alarm clears and returns, the alarm will sound. If a new
alarm occurs the alarm will sound.
8.3 Alarm and Notification Types
All alarms generated are defined as low priority alarms.
The following table contains general descriptions and solutions for the categories of alarms.
Type
Error
Code
Solution
Symbol
Indicator
Light Color
Sound
Indication
Running
Alarms
LXXXX
1. Check if flowmeter is set between 1.0-5.0
L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages (kinked
cannula, blocked vents, etc.)
4. Restart unit using stop/start button
5. Ensure device is in specified operating
environment
If alarm is still present, contact provider
White
with
yellow
dot
Beeping
Shut Down
Alarms
HXXXX
1. Check if flowmeter is set between 1.0-5.0
L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages (kinked
cannula, blocked vents, etc.)
4. Restart unit using stop/start button
5. Ensure device is in specified operating
environment
If alarm is still present, contact provider
White
with red
dot
Constant
Audible
Alarm
Notifications
LXXXX
Contact provider for service
White
with blue
dot
N/A
Invacare® Platinum® 5NXG Oxygen Concentrator