Guidance and manufacturer´s declaration—electromagnetic immunity
The Platinum® 5NXG Concentrator is intended for use in the electromagnetic environment specified below. The
customer or the user of the Platinum® 5NXG Concentrator should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Immunity test
IMMUNITY to
proximity fields
from RF wireless
communications
equipment
MHz –
Modulation –
Field
Strength
385 - 18 Hz - 27
V/m
450 - 18 Hz - 28
V/m
710 - 217 Hz - 9
V/m
745 - 217 Hz - 9
V/m
780 - 217 Hz - 9
V/m
810 - 18 Hz - 28
V/m
870 - 18 Hz - 28
V/m
930 - 18 Hz - 28
V/m
1720 - 217 Hz -
28 V/m
1845 - 217 Hz -
28 V/m
1970 - 217 Hz -
28 V/m
2450 - 217 Hz -
28 V/m
5240 - 217 Hz -
9 V/m
5500 - 217 Hz -
9 V/m
5785 - 217 Hz -
9 V/m
MHz –
Modulation –
Field
Strength
385 - 18 Hz - 27
V/m
450 - 18 Hz - 28
V/m
710 - 217 Hz - 9
V/m
745 - 217 Hz - 9
V/m
780 - 217 Hz - 9
V/m
810 - 18 Hz - 28
V/m
870 - 18 Hz - 28
V/m
930 - 18 Hz - 28
V/m
1720 - 217 Hz -
28 V/m
1845 - 217 Hz -
28 V/m
1970 - 217 Hz -
28 V/m
2450 - 217 Hz -
28 V/m
5240 - 217 Hz - 9
V/m
5500 - 217 Hz - 9
V/m
5785 - 217 Hz - 9
V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the Platinum® 5NXG Concentrator, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
E = [6/d] √P
d = [6/E] √P
where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic
site survey,should be less than the compliance level in
each frequency range.Interference may occur in the
vicinity of equipment marked with the following
symbol:
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.