60126903-A
40
Alarm
Error
Code
Solution
Symbol
Indicator
Light Color
Sound
Indication
Calibration Failure
H7000
1. Restart unit using stop/start
button
2. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Invalid State
H7100
1. Restart unit using stop/start
button
2. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
x = numeric value (defined in the service manual)
Alarm
Error
Code
Solution
Symbol
Indicator
Light Color
Sound
Indication
Notifications
Prediction Notification/Replace
Sieve Bed
L0100
1. Contact provider for sieve
bed replacement
White
with blue
dot
N/A
Prediction Notification/Replace
Compressor or Inlet Filter
L0110
1. Contact provider for
compressor or inlet filter
replacement
White
with blue
dot
N/A
Invacare® Platinum® 5NXG Oxygen Concentrator